Validating pore size estimates in a complex microfiber environment on a human MRI system

Chu Chung Huang, Chih Chin Heather Hsu, Feng Lei Zhou, Slawomir Kusmia, Mark Drakesmith, Geoff J.M. Parker, Ching Po Lin, Derek K. Jones*

*此作品的通信作者

研究成果: Article同行評審

5 引文 斯高帕斯(Scopus)

摘要

Purpose: Recent advances in diffusion-weighted MRI provide “restricted diffusion signal fraction” and restricting pore size estimates. Materials based on co-electrospun oriented hollow cylinders have been introduced to provide validation for such methods. This study extends this work, exploring accuracy and repeatability using an extended acquisition on a 300 mT/m gradient human MRI scanner, in substrates closely mimicking tissue, that is, non-circular cross-sections, intra-voxel fiber crossing, intra-voxel distributions of pore-sizes, and smaller pore-sizes overall. Methods: In a single-blind experiment, diffusion-weighted data were collected from a biomimetic phantom on a 3T Connectom system using multiple gradient directions/diffusion times. Repeated scans established short-term and long-term repeatability. The total scan time (54 min) matched similar protocols used in human studies. The number of distinct fiber populations was estimated using spherical deconvolution, and median pore size estimated through the combination of CHARMED and AxCaliber3D framework. Diffusion-based estimates were compared with measurements derived from scanning electron microscopy. Results: The phantom contained substrates with different orientations, fiber configurations, and pore size distributions. Irrespective of one or two populations within the voxel, the pore-size estimates (~5 μm) and orientation-estimates showed excellent agreement with the median values of pore-size derived from scanning electron microscope and phantom configuration. Measurement repeatability depended on substrate complexity, with lower values seen in samples containing crossing-fibers. Sample-level repeatability was found to be good. Conclusion: While no phantom mimics tissue completely, this study takes a step closer to validating diffusion microstructure measurements for use in vivo by demonstrating the ability to quantify microgeometry in relatively complex configurations.

原文English
頁(從 - 到)1514-1530
頁數17
期刊Magnetic Resonance in Medicine
86
發行號3
DOIs
出版狀態Published - 9月 2021

指紋

深入研究「Validating pore size estimates in a complex microfiber environment on a human MRI system」主題。共同形成了獨特的指紋。

引用此