Thermopneumatic suction integrated microfluidic blood analysis system

Chiao Hsun Yang, Yu Ling Hsieh, Ping Hsien Tsou, Bor-Ran Li*

*此作品的通信作者

研究成果: Article同行評審

21 引文 斯高帕斯(Scopus)

摘要

Blood tests provide crucial diagnostic information regarding several diseases. A key factor that affects the precision and accuracy of blood tests is the interference of red blood cells; however, the conventional methods of blood separation are often complicated and time consuming. In this study, we devised a simple but high-efficiency blood separation system on a self-strained microfluidic device that separates 99.7 ± 0.3% of the plasma in only 6 min. Parameters, such as flow rate, design of the filter trench, and the relative positions of the filter trench and channel, were optimized through microscopic monitoring. Moreover, this air-difference-driven device uses a cost-effective and easy-to-use heater device that creates a low-pressure environment in the microchannel within minutes. With the aforementioned advantages, this blood separation device could be another platform choice for point-of-care testing.

原文English
文章編號e0208676
頁(從 - 到)1-15
頁數15
期刊PLoS ONE
14
發行號3
DOIs
出版狀態Published - 3月 2019