The Effect of Nanoparticle Morphology on the Measurement Accuracy of Mobility Particle Sizers

Amit Awasthi, Bo Sen Wu, Chun Nan Liu, Chun Wan Chen, Shi Nan Uang, Chuen-Tinn Tsai*

*此作品的通信作者

研究成果: Article同行評審

16 引文 斯高帕斯(Scopus)

摘要

The influence of particle morphology on the accuracy of nanoparticle size distributions measured by the engine exhaust particle sizer spectrometer (EEPS, TSI Model 3090) was studied using the scanning mobility particle sizer (SMPS, TSI Model 3936) as a reference. The EEPS shows higher total number concentrations with the maximum relative difference up to 67 % and smaller number median mobility diameters for polydisperse silver nanoparticles generated in the laboratory. To provide a quantitative explanation of the difference, generated polydisperse nanoparticles were classified as monodisperse particles with the initial equivalent mobility diameter (dm1) and sintered in the second furnace at different temperatures (room temperature to 600 °C), to change their morphologies for the comparison tests. Without sintering (room temperature), results show that the measured mobility diameter (dm2) of the EEPS is smaller than that measured by the SMPS when dm1 is larger than 30 nm and the difference increases as dm1 is increased from 30 to 300 nm. But the difference decreases as the morphology of particles is changed from branched chain agglomerates to spheres for dm1 less than 80 nm and the sintering temperature higher than 200 °C. Theoretical analysis shows that the mean charge per agglomerates is more than that of spheres resulting in overestimation of the electrical mobility and underestimation of the dm2 by the EEPS.

原文English
頁(從 - 到)205-215
頁數11
期刊Mapan - Journal of Metrology Society of India
28
發行號3
DOIs
出版狀態Published - 9月 2013

指紋

深入研究「The Effect of Nanoparticle Morphology on the Measurement Accuracy of Mobility Particle Sizers」主題。共同形成了獨特的指紋。

引用此