Clinical ultrasonic systems provide only video output signals which limit the data acquisition and signal processing capability from current systems. The purpose of this study is to develop a pulsed Doppler ultrasound signal processing unit. The architecture of this system include: central frequency generator, PRF generator, echo signal demodulator, and sample volume setting circuit. The computation architecture included FFT Doppler spectrum, autocorrelation phase shift and cross-correlation time shift analysis. The acquired RF and baseband signals were used for calculations of FFT, autocorrelation, and cross-correlation to verify the flow velocity measurements.