TY - JOUR
T1 - Tailor-made iliac branched device for preserving the internal iliac artery in patients with common iliac artery aneurysm
AU - Chen, Po Lin
AU - Hsu, Hung Lung
AU - Chen, I. Ming
AU - Kuo, Tzu Ting
AU - Chen, Yin Yin
AU - Shih, Chun Che
N1 - Publisher Copyright:
Copyright © 2019, the Chinese Medical Association.
PY - 2019
Y1 - 2019
N2 - Background: Endovascular repair with stent-graft is a treatment option for patient with common iliac artery aneurysm (CIAA). However, the preservation of the internal iliac artery (IIA) is a concern. The commercially available iliac branched device (IBD) requires a common iliac length of at least 5cm, which is usually too long for Asian people. Here, we report our medium-term results of using tailor-made IBD for patients with short common iliac artery (CIA) with and without abdominal aortic aneurysm (AAA). Methods: A selected iliac limb of the AAA stent-graft was unloaded from the delivery system. A 6-mm fenestration hole was made at the length of the CIA from the proximal end. The edge of the hole was reinforced with the soft and radiopaque tip of a 0.014´´ wire. Then, the iliac limb was reloaded into the introduced sheath as the tailor-made IBD. It was inserted from the selected side of the femoral artery and deployed. The ipsilateral IIA was cannulated through the fenestration hole. Then, a balloon-expandable or self-expandable covered stent with an appropriate size was deployed as the bridging stent-graft. Results: Between March 2013 and March 2017, a total of 10 patients received the tailor-made IBDs. One patient died of systemic thromboembolism 2 days after the operation. The bridging stent-grafts remained patent in all patients, except one occluded at 1 year after operation. Conclusion: A tailor-made IBD is an easy-to-apply, alternative option for preserving the IIA perfusion in short CIAA patients with and without AAA.
AB - Background: Endovascular repair with stent-graft is a treatment option for patient with common iliac artery aneurysm (CIAA). However, the preservation of the internal iliac artery (IIA) is a concern. The commercially available iliac branched device (IBD) requires a common iliac length of at least 5cm, which is usually too long for Asian people. Here, we report our medium-term results of using tailor-made IBD for patients with short common iliac artery (CIA) with and without abdominal aortic aneurysm (AAA). Methods: A selected iliac limb of the AAA stent-graft was unloaded from the delivery system. A 6-mm fenestration hole was made at the length of the CIA from the proximal end. The edge of the hole was reinforced with the soft and radiopaque tip of a 0.014´´ wire. Then, the iliac limb was reloaded into the introduced sheath as the tailor-made IBD. It was inserted from the selected side of the femoral artery and deployed. The ipsilateral IIA was cannulated through the fenestration hole. Then, a balloon-expandable or self-expandable covered stent with an appropriate size was deployed as the bridging stent-graft. Results: Between March 2013 and March 2017, a total of 10 patients received the tailor-made IBDs. One patient died of systemic thromboembolism 2 days after the operation. The bridging stent-grafts remained patent in all patients, except one occluded at 1 year after operation. Conclusion: A tailor-made IBD is an easy-to-apply, alternative option for preserving the IIA perfusion in short CIAA patients with and without AAA.
KW - Abdominal aortic aneurysm
KW - Common iliac artery aneurysm
KW - Endovascular aneurysm repair
KW - Iliac branched device
KW - Internal iliac artery
UR - http://www.scopus.com/inward/record.url?scp=85071783743&partnerID=8YFLogxK
U2 - 10.1097/JCMA.0000000000000156
DO - 10.1097/JCMA.0000000000000156
M3 - Article
C2 - 31335630
AN - SCOPUS:85071783743
SN - 1726-4901
VL - 82
SP - 710
EP - 713
JO - Journal of the Chinese Medical Association
JF - Journal of the Chinese Medical Association
IS - 9
ER -