Prediction of residual astigmatism in cataract surgery at different diameter zones using optical biometry measurement

Yin Hsi Chang, Christy Pu, Ken Kuo Lin, Jiahn Shing Lee, Chiun Ho Hou*

*此作品的通信作者

研究成果: Article同行評審

3 引文 斯高帕斯(Scopus)

摘要

The studies for astigmatism prediction error at different diameters using optical biometry are scant. We investigated patients who underwent cataract surgery with monofocal, nontoric intraocular lens (IOL) from 2017 through 2019 in a medical center. Patients with prior refractive surgeries, corneal opacity, or surgical complications were excluded. Corneal astigmatism (CA) was measured using AL-Scan at 2.4- and 3.3-mm diameter zones and calculated using the Barrett toric calculator preoperatively and postoperatively. The mean absolute error and centroid prediction error for the two zones were computed using double-angle plots. In total, 101 eyes of 76 patients were analyzed. Mean patient age was 68.7 ± 9.3 years and mean preoperative CA power was 0.7 ± 0.5 D. The overall centroid prediction error a 3.3 mm (0.09 ± 0.58 D@25) was significantly lower than that at 2.4 mm (0.09 ± 0.68 D@87) on the X-axis (P = 0.003). The 3.3-mm measurement also had a lower centroid prediction error than the 2.4-mm did for eyes with against-the-rule (ATR) and oblique astigmatism (P = 0.024; 0.002 on X-axis, respectively). The 3.3-mm measurement provided a more accurate CA estimation than the 2.4-mm did, particularly for ATR astigmatism. Diameter zone and astigmatism type should be considered crucial to precise astigmatism calculation.

原文English
文章編號4305
期刊Scientific reports
12
發行號1
DOIs
出版狀態Published - 12月 2022

指紋

深入研究「Prediction of residual astigmatism in cataract surgery at different diameter zones using optical biometry measurement」主題。共同形成了獨特的指紋。

引用此