Office blood pressure measurement: A comprehensive review

Saulat Siddique*, Aamir Hameed Khan, Hunaina Shahab, Yu Qing Zhang, Jam Chin Tay, Peera Buranakitjaroen, Yuda Turana, Narsingh Verma, Chen Huan Chen, Hao Min Cheng, Tzung Dau Wang, Huynh Van Minh, Yook Chin Chia, Kazuomi Kario

*此作品的通信作者

研究成果: Review article同行評審

12 引文 斯高帕斯(Scopus)

摘要

The conventional auscultatory methods for measuring blood pressure have been used to screen, diagnose, and manage hypertension since long. However, these have been found to be prone to errors especially the white coat phenomena which cause falsely high blood pressure readings. The Mercury sphygmomanometer and the Aneroid variety are no longer recommended by WHO for varying reasons. The Oscillometric devices are now recommended with preference for the Automated Office Blood Pressure measurement device which was found to have readings nearest to the Awake Ambulatory Blood Pressure readings. The downside for this device is the cost barrier. The alternative is to use the simple oscillometric device, which is much cheaper, with the rest and isolation criteria of the SPRINT study. This too may be difficult due to space constraints and the post-clinic blood measurement is a new concept worth further exploration.

原文English
頁(從 - 到)440-449
頁數10
期刊Journal of Clinical Hypertension
23
發行號3
DOIs
出版狀態Published - 3月 2021

指紋

深入研究「Office blood pressure measurement: A comprehensive review」主題。共同形成了獨特的指紋。

引用此