Noninvasive glucose monitoring in vivo with an optical heterodyne polarimeter

Chien Chou*, Chien Yuan Han, Wen Chuan Kuo, Yeu Chuen Huang, Ching Mei Feng, Jenn Chyang Shyu

*此作品的通信作者

研究成果: Article同行評審

89 引文 斯高帕斯(Scopus)

摘要

An amplitude-sensitive optical heterodyne polarimeter was set up to monitor noninvasively the aqueous glucose concentration in a rabbit's eye. A Zeeman laser in conjunction with a Clan-Thompson analyzer was used to generate an optical heterodyne signal. The amplitude of the heterodyne signal linearly related to the optical rotation angle of the aqueous glucose. The concentration of the aqueous glucose in a rabbit's eyeball was measured in vivo. There was a 30-min time delay between observations of aqueous glucose and blood glucose. The detection capability and the reproducibility of the experiment are demonstrated and discussed.

原文English
頁(從 - 到)3553-3557
頁數5
期刊Applied Optics
37
發行號16
DOIs
出版狀態Published - 1998

指紋

深入研究「Noninvasive glucose monitoring in vivo with an optical heterodyne polarimeter」主題。共同形成了獨特的指紋。

引用此