Noninvasive blood pressure analysis using Empirical Mode Decomposition (EMD) - Based oscillometric method

Fu Jung Hsiao*, Chen Chuan Hung

*此作品的通信作者

研究成果: Conference contribution同行評審

4 引文 斯高帕斯(Scopus)

摘要

This study proposes a new method to extract oscillometric pulses for noninvasive blood pressure measurement using empirical mode decomposition, based on the oscillometric method. The mean and standard deviation of the present method in systolic and diastolic pressure are 1.6 and 5.2, and 0.7 and 5.2, respectively, which not only passes the AAMI (Association for the Advancement of Medical Instrumentation) and EN (European Norm) blood pressure measurement criteria, but also performs more accurate measurements than commercial noninvasive blood pressure measurement devices.

原文English
主出版物標題Proceedings - 2010 3rd International Conference on Biomedical Engineering and Informatics, BMEI 2010
頁面973-977
頁數5