Long-term evaluation and calibration of three types of low-cost PM2.5 sensors at different air quality monitoring stations

Gung Hwa Hong, Thi Cuc Le, Jing Wei Tu, Chieh Wang, Shuenn Chin Chang, Jhih Yuan Yu, Guan Yu Lin, Shankar G. Aggarwal, Chuen-Jinn Tsai*

*此作品的通信作者

研究成果: Article同行評審

8 引文 斯高帕斯(Scopus)

摘要

To evaluate the performance of low-cost PM2.5 sensors and develop calibration models for correcting for the PM2.5 sensor data (PM2.5,S), field comparison tests were conducted based on Met One BAM-1020 data at various locations using long-term (≥one year) data of Plantower PMS5003, Sensirion SPS30, and Honeywell HPMA115S0 PM2.5 sensors. Both multivariate linear regression (MLR) and non-linear regression (NLR) models using hourly RHs and original sensor PM2.5 data as parameters were able to obtain accurate calibrated hourly PM2.5 values with MNBs (mean normalized biases) less than about ±10% and MNEs (mean normalized errors) less than about 30% for all three types of PM2.5 sensors at all monitoring locations. On the other hand, the MNB and MNE of the calibrated 24-hr average PM2.5 data for the two models were less than ±13% and 20%, respectively. Moreover, the slope, intercept, and R2 of the linear regression line of the calibrated 24-hr average PM2.5 and BAM-1020 data were as good as 1.0 ± 0.1, 0.0 ± 2.0 μg/m3, and ≥0.88, respectively. Therefore, these well-calibrated sensors can well be served for education and information (MNE<50%), hotspot identification and characterization (MNE<30%), and personal exposure study (MNE<30%) purposes, and even supplement the existing daily PM2.5 data of the air quality monitoring stations (MNE<20%).

原文English
文章編號105829
頁(從 - 到)1-16
頁數16
期刊Journal of Aerosol Science
157
DOIs
出版狀態Published - 9月 2021

指紋

深入研究「Long-term evaluation and calibration of three types of low-cost PM2.5 sensors at different air quality monitoring stations」主題。共同形成了獨特的指紋。

引用此