Leadless pacemaker implementation at the right atrial appendage apex: An initial preclinical assessment

Yu Sheng Lin, Lung Sheng Wu, Wan Chun Ho, Chao Sung Lai, Wilber Su, Pao Hsien Chu*

*此作品的通信作者

研究成果: Article同行評審

摘要

Objective: This study evaluates the feasibility and efficacy of implanting a leadless pacemaker at the right atrial appendage (RAA) in a preclinical minipig model, aiming to address the limitations of atrial pacing with current leadless devices like the Medtronic Micra, which is typically used for right ventricular implantation. Methods: Four minipigs, each with a median body weight of 45.8 ± 10.0 kg, underwent placement of the Micra transcatheter pacing system (TPS) via the right femoral vein into the RAA apex. The pacing performance was assessed over 1-week (short-term) and 3-month (long-term) periods. Outcomes: The initial findings indicated successful implantation, with satisfactory intrinsic R-wave amplitudes and pacing threshold. In the following period, the sensitivity, threshold, and impedance were stable with time. Notably, upon explanation at 3 months, a deep myocardial penetration by the device was observed, necessitating a redesign for safe long-term use in a growing subject's heart. Conclusion: While initial results suggest that RAA apex placement of the Micra TPS is promising for potential inclusion in a dual-chamber pacing system, the issue of myocardial penetration highlights the need for device redesign to ensure safety and effectiveness in long-term applications.

原文English
頁(從 - 到)1124-1127
頁數4
期刊PACE - Pacing and Clinical Electrophysiology
47
發行號8
DOIs
出版狀態Published - 8月 2024

指紋

深入研究「Leadless pacemaker implementation at the right atrial appendage apex: An initial preclinical assessment」主題。共同形成了獨特的指紋。

引用此