Integration of a guided-mode resonance filter with microposts for in-cell protein detection

Yi Kai Tu, Meng Zhe Tsai, I. Chin Lee, Hsin-Yun Hsu, Cheng-Sheng Huang*

*此作品的通信作者

研究成果: Article同行評審

25 引文 斯高帕斯(Scopus)

摘要

We present an integrated microfluidic system consisting of a label-free biosensor of a guided-mode resonance filter (GMRF) and a microfluidic channel with a micropost filter. The GMRF was fabricated through replica molding using an ultraviolet-curable polymer and a plastic substrate. An array of microposts (a diameter and height of 26.5 and 56 μm, respectively, and a spacing between 7.5 and 9.5 μm), fabricated on a silicon substrate through photolithography, was used as the filter. A double-sided tape was used to laminate the GMRF and a microfluidic chip such that the integrated device provides two functions: filtration of the cell debris and quantification of the in-cell protein concentration. By measuring the changes in the resonant wavelength from the GMRF, the detection of β-actin in an unprocessed lysed cell sample was demonstrated; the cell debris was separated using the micropost filter to prevent false measurement.

原文English
頁(從 - 到)4189-4195
頁數7
期刊Analyst
141
發行號13
DOIs
出版狀態Published - 29 4月 2016

指紋

深入研究「Integration of a guided-mode resonance filter with microposts for in-cell protein detection」主題。共同形成了獨特的指紋。

引用此