Finite element analysis after rod fracture of the spinal hybrid elastic rod system

Jui Yang Hsieh, Chen Sheng Chen, Shao Ming Chuang, Jyh Horng Wang, Po Quang Chen, Yi You Huang*

*此作品的通信作者

研究成果: Article同行評審

2 引文 斯高帕斯(Scopus)

摘要

Background: The spinal hybrid elastic (SHE) rod dynamic stabilization system can provide sufficient spine support and less adjacent segment stress. This study aimed to investigate the biomechanical effects after the internal fracture of SHE rods using finite element analysis. Methods: A three-dimensional nonlinear finite element model was developed. The SHE rod comprises an inner nitinol stick (NS) and an outer polycarbonate urethane (PCU) shell (PS). The fracture was set at the caudal third portion of the NS, where the maximum stress occurred. The resultant intervertebral range of motion (ROM), intervertebral disc stress, facet joint contact force, screw stress, NS stress, and PCU stress were analyzed. Results: When compared with the intact spine model, the overall trend was that the ROM, intervertebral disc stress, and facet joint force decreased in the implanted level and increased in the adjacent level. When compared with the Ns-I, the trend in the Ns-F decreased and remained nearly half effect. Except for torsion, the PS stress of the Ns-F increased because of the sharing of NS stress after the NS fracture. Conclusions: The study concluded the biomechanical effects still afford nearly sufficient spine support and gentle adjacent segment stress after rod fracture in a worst-case scenario of the thinnest PS of the SHE rod system.

原文English
文章編號816
期刊BMC Musculoskeletal Disorders
23
發行號1
DOIs
出版狀態Published - 12月 2022

指紋

深入研究「Finite element analysis after rod fracture of the spinal hybrid elastic rod system」主題。共同形成了獨特的指紋。

引用此