Efficacy of a computerized sensor system for evaluation and training of dizzy patients

Chung Lan Kao*, Wan Ling Hsieh, Shuu Jiun Wang, Shih Jen Chen, Shun Hwa Wei, Rai Chi Chan

*此作品的通信作者

研究成果: Article同行評審

4 引文 斯高帕斯(Scopus)

摘要

Patients with vestibular hypofunction often experience dizziness and unsteadiness while moving their heads. Appropriate sensors can effectively detect a patient's dynamic visual acuity and associated body balance control. Forty-one vestibular-deficit patients and 10 normal individuals were invited to participate in this study. Questionnaires, clinical assessment scales and objective measures were evaluated on participants' first visits. After 12 sessions of training, all scales were evaluated again on vestibular-deficit patients. The computerized system was composed of sensors, including a gyro and strain gauges, data acquisition accessories and LabVIEW software. Results revealed that the system could effectively distinguish normal subjects from subjects with vestibular deficits. In addition, after a rehabilitation program, subjects' subjective and objective performances were significantly improved. Based on our results, we concluded that the present system, which uses a gyro and strain gauges, may provide an effective method for assessing and treating vestibular-deficit patients.

原文English
頁(從 - 到)7602-7620
頁數19
期刊Sensors
10
發行號8
DOIs
出版狀態Published - 8月 2010

指紋

深入研究「Efficacy of a computerized sensor system for evaluation and training of dizzy patients」主題。共同形成了獨特的指紋。

引用此