Effect of cutting flute design features on primary stability of immediate implant placement and restoration: a dynamic experimental analysis

  • Min Chieh Hsieh
  • , Chang Hung Huang
  • , Ming Lun Hsu*
  • *此作品的通信作者

研究成果: Article同行評審

5 引文 斯高帕斯(Scopus)

摘要

Self-tapping implants with self-cutting flutes may influence primary stability, especially for the immediate implant placement and restoration protocol in which implants are affixed to the bone in the apical portion. Screw geometry differs between brands, and the effect of apical design on its clinical outcomes remains unclear. This study is aimed at investigating the influence of cutting flute shape (spiral, straight, and without flute) on primary stability by using a dynamic experimental test. Six types of dental implants were designed using computer-aided design and computer-aided manufacturing technology, consisting of three types of cutting flute shapes along with two types of screw features. A dynamic mechanical test was performed using a cyclic loading scheme. The mechanical behaviors of resistance to lateral load (RLL), maximum force, and energy dissipation were compared between groups. In the dynamic test, implants without cutting flute also exhibited higher values in RLL, maximum force, and energy dissipation. The aggressive thread implant with straight flute displayed higher RLL and had a significantly higher values in RLL (p = 0.033) at the threshold point of bone–implant interface breakdown. The implants without cutting flutes exhibited higher primary stability. Straight flute design would improve RLL for aggressive thread implant. Graphical Abstract: [Figure not available: see fulltext.]

原文English
頁(從 - 到)475-484
頁數10
期刊Medical and Biological Engineering and Computing
61
發行號2
DOIs
出版狀態Published - 2月 2023

指紋

深入研究「Effect of cutting flute design features on primary stability of immediate implant placement and restoration: a dynamic experimental analysis」主題。共同形成了獨特的指紋。

引用此