Doppler angle and flow velocity estimations using the classic and transverse doppler effects

Bor Ray Lee*, Huihua Kenny Chiang, Cheng Deng Kuo, Win Li Lin, San Kern Lee

*此作品的通信作者

研究成果: Article同行評審

21 引文 斯高帕斯(Scopus)

摘要

Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, arid the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, we developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer lo estimate the Doppler angle and the flow velocity. We have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45° and 80° Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5°. This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity.

原文English
頁(從 - 到)252-256
頁數5
期刊IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
46
發行號1
DOIs
出版狀態Published - 1999

指紋

深入研究「Doppler angle and flow velocity estimations using the classic and transverse doppler effects」主題。共同形成了獨特的指紋。

引用此