摘要
This work develops a label-free gliadin immunosensor that is based on changes in the frequency of a quartz crystal microbalance (QCM) chip. A higher sensitivity was obtained by applying 25 nm gold nanoparticles (AuNPs) to the surface of a bare QCM electrode. Subsequently, chicken anti-gliadin antibodies (IgY) were immobilized directly on the AuNP-modified surface by cross-linking amine groups in IgY with glutaraldehyde. Experimental results revealed that the change in frequency exhibited when 2 ppm gliadin was bound to the AuNP-modified electrode was 35 Hz (48%) greater than that of the bare gold electrode. The linear dynamic range in 60% ethanol was from 1 × 101 to 2 × 105 ppb gliadin, and the calculated limit of detection (LOD) was 8 ppb. The entire detection process was completed in 40 min and was highly repeatable. Additionally, the AuNP-modified QCM system generated results in the detection of gliadin in 10 commercial food products that were consistent with those obtained using an AOAC-approved gliadin kit. In conclusion, the QCM platform provides a potential alternative means of ensuring that people with wheat allergies and celiac patients have access to gliadin-free food.
原文 | English |
---|---|
頁(從 - 到) | 6483-6492 |
頁數 | 10 |
期刊 | Journal of Agricultural and Food Chemistry |
卷 | 60 |
發行號 | 26 |
DOIs | |
出版狀態 | Published - 4 7月 2012 |