Biomechanical consideration of total hip arthroplasty following failed fixation of femoral intertrochanteric fractures - A finite element analysis

Dave W. Chen, Chun Li Lin, Chih Chien Hu, Ming Feng Tsai, Mel S. Lee*

*此作品的通信作者

研究成果: Article同行評審

22 引文 斯高帕斯(Scopus)

摘要

Objective: The current study aimed to perform a biomechanical analysis of the effect and distribution of stress and strain values in the femur and its residual screw holes from the implantation of different lengths of total hip prostheses following failed dynamic hip screw (DHS) fixation of AO type A1.1 and A1.3 intertrochanteric fractures. Methods: Medical image processing, computer-assisted engineering design, and finite element analysis were combined for the analysis and the reliability of the model confirmed by convergence testing and comparison with a physical model. Using finite element analysis, patterns of stress from implantation of total hip prostheses with different stem length were obtained. Results: Stress distribution was concentrated over the third and fourth residual screw holes, especially in the models with shorter stem length. Fracture type (AO type A1.1 or A1.3) did not alter its distribution. In proximal cancellous bone, a stress-shielding effect was seen at the region of the residual fracture fragment, especially in the A1.1 fracture. In the fracture fixed with 2 cable wires, the maximum von Mises stress was in the proximal cable wire. Conclusion: The results suggest that an increase in the original stem length equal to the diameter of the femoral isthmus, or a distance between the most distal residual screw hole and the end of the femoral prosthesis, provide improved stress distribution.

原文English
頁(從 - 到)569-575
頁數7
期刊Medical Engineering and Physics
35
發行號5
DOIs
出版狀態Published - 5月 2013

指紋

深入研究「Biomechanical consideration of total hip arthroplasty following failed fixation of femoral intertrochanteric fractures - A finite element analysis」主題。共同形成了獨特的指紋。

引用此