Auto-focus pathology microscope using sub-array sampling

Po Hsun Wu, Chun Jen Weng, Pi Ying Cheng

    研究成果: Conference contribution同行評審

    1 引文 斯高帕斯(Scopus)

    摘要

    this study compared the focusing results of binning and sub-array methods with the aim of balancing speed and accuracy. Analysis was conducted using an autofocus system comprising a microscope, CCD, PZT actuator, LED light source, and two focus algorithms. Our results showed that by selecting an image sub-array of appropriate size and shape, it was possible to reduce the time required for autofocus operations to only 1 second, while maintaining the focusing accuracy of full-field analysis techniques.

    原文English
    主出版物標題I2MTC 2019 - 2019 IEEE International Instrumentation and Measurement Technology Conference, Proceedings
    發行者Institute of Electrical and Electronics Engineers Inc.
    ISBN(電子)9781538634608
    DOIs
    出版狀態Published - 五月 2019
    事件2019 IEEE International Instrumentation and Measurement Technology Conference, I2MTC 2019 - Auckland, New Zealand
    持續時間: 20 五月 201923 五月 2019

    出版系列

    名字Conference Record - IEEE Instrumentation and Measurement Technology Conference
    2019-May
    ISSN(列印)1091-5281

    Conference

    Conference2019 IEEE International Instrumentation and Measurement Technology Conference, I2MTC 2019
    國家/地區New Zealand
    城市Auckland
    期間20/05/1923/05/19

    指紋

    深入研究「Auto-focus pathology microscope using sub-array sampling」主題。共同形成了獨特的指紋。

    引用此