TY - JOUR
T1 - Application of an interspinous process device after minimally invasive lumbar decompression could lead to stress redistribution at the pars interarticularis
T2 - A finite element analysis
AU - Lo, Hao Ju
AU - Chen, Chen Sheng
AU - Chen, Hung Ming
AU - Yang, Sai Wei
N1 - Publisher Copyright:
© 2019 The Author(s).
PY - 2019/5/15
Y1 - 2019/5/15
N2 - Background: An interspinous process device, the Device for Intervertebral Assisted Motion (DIAM™) designed to treat lumbar neurogenic disease secondary to the lumbar spinal stenosis, it provides dynamic stabilization after minimally invasive (MI) lumbar decompression. The current study was conducted using an experimentally validated L1-L5 spinal finite element model (FEM) to evaluate the limited decompression on range of motion (ROM) and stress distribution on a neural arch implanted with the DIAM. Methods: The study simulated bilateral laminotomies with partial discectomy at L3-L4, as well as unilateral and bilateral laminotomies with partial discectomy combined with implementation of the DIAM at L3-L4. The ROM and maximum von Mises stresses in flexion, extension, lateral bending, and axial torsion were analyzed in response to the hybrid protocol in comparison with the intact model. Results: The investigation revealed that decreased ROM, intradiscal stress, and facet joint force at the implant level, but considerably increased stress at the pars interarticularis were found during flexion and torsion at the L4, as well as during extension, lateral bending, and torsion at the L3, when the DIAM was implanted compared with the defect model. Conclusion: The results demonstrate that the DIAM may be beneficial in reducing the symptoms of stress-induced low back pain. Nevertheless, the results also suggest that a surgeon should be cognizant of the stress redistribution at the pars interarticularis results from MI decompression plus the application of the interspinous process device.
AB - Background: An interspinous process device, the Device for Intervertebral Assisted Motion (DIAM™) designed to treat lumbar neurogenic disease secondary to the lumbar spinal stenosis, it provides dynamic stabilization after minimally invasive (MI) lumbar decompression. The current study was conducted using an experimentally validated L1-L5 spinal finite element model (FEM) to evaluate the limited decompression on range of motion (ROM) and stress distribution on a neural arch implanted with the DIAM. Methods: The study simulated bilateral laminotomies with partial discectomy at L3-L4, as well as unilateral and bilateral laminotomies with partial discectomy combined with implementation of the DIAM at L3-L4. The ROM and maximum von Mises stresses in flexion, extension, lateral bending, and axial torsion were analyzed in response to the hybrid protocol in comparison with the intact model. Results: The investigation revealed that decreased ROM, intradiscal stress, and facet joint force at the implant level, but considerably increased stress at the pars interarticularis were found during flexion and torsion at the L4, as well as during extension, lateral bending, and torsion at the L3, when the DIAM was implanted compared with the defect model. Conclusion: The results demonstrate that the DIAM may be beneficial in reducing the symptoms of stress-induced low back pain. Nevertheless, the results also suggest that a surgeon should be cognizant of the stress redistribution at the pars interarticularis results from MI decompression plus the application of the interspinous process device.
KW - Finite element model
KW - Interspinous process device
KW - Minimally invasive
KW - Pars interarticularis
UR - http://www.scopus.com/inward/record.url?scp=85066041692&partnerID=8YFLogxK
U2 - 10.1186/s12891-019-2565-5
DO - 10.1186/s12891-019-2565-5
M3 - Article
C2 - 31092237
AN - SCOPUS:85066041692
SN - 1471-2474
VL - 20
JO - BMC Musculoskeletal Disorders
JF - BMC Musculoskeletal Disorders
IS - 1
M1 - 213
ER -