A time-delay calibrated method for cornea hysteresis and intraocular pressure measurement

Kuo Jen Wang, Che Liang Tsai*, Wai Wang, Long Hsu, Ken-Yuh Hsu

*此作品的通信作者

研究成果: Article同行評審

摘要

The presence of cornea hysteresis (CH) in characterizing the intraocular pressure (IOP) of a human eye deteriorates the accuracy of IOP. To suppress CH, the pressure gauge of a tonometer must be located as close as possible to the cornea. However, this arrangement is unpractical because appropriate working distance to the cornea is required. In this paper, a time-delay calibrated (TDC) method is proposed to counteract the undesired effect of CH in characterizing the IOP. Employing this TDC method, the CH approaches to zero for most eyes measured.