Mathematical derivation, circuits design and clinical experiments of measuring blood flow volume (BFV) at arteriovenous fistula (AVF) of hemodialysis (HD) patients using a newly-developed photoplethysmography (PPG) sensor

Paul C.P. Paul, Pei Yu Chiang, Der-Cherng Tarng, C. Y. Yang

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard.

Original languageEnglish
Title of host publication2018 19th International Symposium on Quality Electronic Design, ISQED 2018
PublisherIEEE Computer Society
Pages353-356
Number of pages4
ISBN (Electronic)9781538612149
DOIs
StatePublished - 9 May 2018
Event19th International Symposium on Quality Electronic Design, ISQED 2018 - Santa Clara, United States
Duration: 13 Mar 201814 Mar 2018

Publication series

NameProceedings - International Symposium on Quality Electronic Design, ISQED
Volume2018-March
ISSN (Print)1948-3287
ISSN (Electronic)1948-3295

Conference

Conference19th International Symposium on Quality Electronic Design, ISQED 2018
Country/TerritoryUnited States
CitySanta Clara
Period13/03/1814/03/18

Keywords

  • Arteriovenous Fistula (AVF)
  • Blood Flow Volume (BFV)
  • Photoplethysmography (PPG) Sensor