Finite element modelling of implant designs and cortical bone thickness on stress distribution in maxillary type IV bone

Chou I-Chiang, Lee Shyh-Yuan, Wu Ming-Chang, Chia Wei Sun, Cho Pei Jiang*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

42 Scopus citations

Abstract

The aims of this study were to examine the effect of implant neck design and cortical bone thickness using 3D finite element analysis and to analyse the stability of clinical evidence based on micromotion and principal stress. Four commercial dental implants for a type IV bone and maxillary segments were created. Various parameters were considered, including the osseointegration condition, loading direction and cortical bone thickness. Micromotion and principal stresses were used to evaluate the failure of osseointegration and bone overloading, respectively. It was found that the maximum stress of the peri-implant bone decreased as cortical bone thickness increased. The micromotion level in full osseointegration is less than that in non-osseointegration and it also decreases as cortical bone thickness increases. The cortical bone thickness should be measured before surgery to help select a proper implant. In the early stage of implantation, the horizontal loading component induces stress concentration in bone around the implant neck more easily than does the vertical loading component, and this may result in crestal bone loss.

Original languageEnglish
Pages (from-to)516-526
Number of pages11
JournalComputer Methods in Biomechanics and Biomedical Engineering
Volume17
Issue number5
DOIs
StatePublished - 2014

Keywords

  • dental implant
  • maxillary
  • micromotion
  • neck design

Fingerprint

Dive into the research topics of 'Finite element modelling of implant designs and cortical bone thickness on stress distribution in maxillary type IV bone'. Together they form a unique fingerprint.

Cite this