TY - JOUR
T1 - Electrostatic charge measurement and charge neutralization of fine aerosol particles during the generation process
AU - Tsai, Chuen Jinn
AU - Lin, Jyh Shyan
AU - Deshpande, C. G.
AU - Liu, Li Chun
PY - 2006/4/1
Y1 - 2006/4/1
N2 - An aerosol charge analyzer has been constructed to measure the charge distribution of NaCl particles generated in the laboratory. A radioactive electrostatic charge neutralizer utilizing Po-210 was used to neutralize the electrostatic charge of the particles. The atomization technique was used to generate NaCl particles with diameters of 0.2 to 0.8 μm, while the evaporation and condensation method was adopted to generate particles of 0.01 to 0.2 μm in diameter. The experimental data demonstrates that the absolute average particle charge depends on the particle diameter, and is higher than that calculated by the Boltzmann charge equilibrium for particles within the range of 0.2 to 0.8 μm. The charge increases with decreasing NaCl concentration. When these particles are neutralized using the Po-210 neutralizer, it is found that the electrostatic charge reaches the Boltzmann charge equilibrium. For 0.01 to 0.2 μm NaCl particles generated using the evaporation and condensation method, test results show that the absolute average particle charge is higher than that calculated by the Boltzmann charge equilibrium for particles larger than 0.03 to 0.05 μm in diameter, while it is lower than that predicted by the Fuchs theory [1], for particles smaller than 0.03 to 0.05 μm. However, after charge neutralization, particles with diameter above 0.05 μm reach the Boltzmann charge equilibrium condition, and the charges for particles with diameters of 0.010 to 0.05 μm, agree well with Fuchs' theory.
AB - An aerosol charge analyzer has been constructed to measure the charge distribution of NaCl particles generated in the laboratory. A radioactive electrostatic charge neutralizer utilizing Po-210 was used to neutralize the electrostatic charge of the particles. The atomization technique was used to generate NaCl particles with diameters of 0.2 to 0.8 μm, while the evaporation and condensation method was adopted to generate particles of 0.01 to 0.2 μm in diameter. The experimental data demonstrates that the absolute average particle charge depends on the particle diameter, and is higher than that calculated by the Boltzmann charge equilibrium for particles within the range of 0.2 to 0.8 μm. The charge increases with decreasing NaCl concentration. When these particles are neutralized using the Po-210 neutralizer, it is found that the electrostatic charge reaches the Boltzmann charge equilibrium. For 0.01 to 0.2 μm NaCl particles generated using the evaporation and condensation method, test results show that the absolute average particle charge is higher than that calculated by the Boltzmann charge equilibrium for particles larger than 0.03 to 0.05 μm in diameter, while it is lower than that predicted by the Fuchs theory [1], for particles smaller than 0.03 to 0.05 μm. However, after charge neutralization, particles with diameter above 0.05 μm reach the Boltzmann charge equilibrium condition, and the charges for particles with diameters of 0.010 to 0.05 μm, agree well with Fuchs' theory.
KW - Aerosol charge analyzer
KW - Charge equilibrium
KW - Charge neutralizer
KW - Condensation aerosols
KW - Electrostatic charge
KW - Evaporation aerosols
UR - http://www.scopus.com/inward/record.url?scp=33646432221&partnerID=8YFLogxK
U2 - 10.1002/ppsc.200500961
DO - 10.1002/ppsc.200500961
M3 - Article
AN - SCOPUS:33646432221
SN - 0934-0866
VL - 22
SP - 293
EP - 298
JO - Particle and Particle Systems Characterization
JF - Particle and Particle Systems Characterization
IS - 5
ER -