Abstract
In this study, the use and advantages of balanced detection (BD) in spectral domain optical coherence tomography (SD-OCT) are demonstrated. A-scans are calculated as a combination of two phase-opposed interferometric spectra acquired simultaneously by using a multiline single camera spectrometer. Not only does this system suppress artifacts due to autocorrelation, but also the signal of interest is increased by a factor of 2 as experimentally verified. Our BD-based SD-OCT gives a signal-to-noise ratio improvement of 8-14 dB for the peak within 1 mm compared to standard SD-OCT using a single detection scheme. This method is validated by experimental measurement of a glass plate.
Original language | English |
---|---|
Pages (from-to) | 5936-5940 |
Number of pages | 5 |
Journal | Applied Optics |
Volume | 51 |
Issue number | 24 |
DOIs | |
State | Published - 20 Aug 2012 |