Abstract
This paper proposes to use a new and realistic finite element model of a single turn human cochlea and cochlear implant electrode array with electrode-tissue interface to model the electric field imaging technique, which creates an impedance matrix. Due to discrepancy in the Warburg capacitor in the active and nonactive electrode contacts and electric power lost along the cochlea, the diagonal terms of the impedance matrix is larger than the off diagonal terms, which represent the impedance values of the nonactive electrode contacts. The impedance matrix from electric field imaging measurements is compared with the simulation results.
Original language | English |
---|---|
Pages (from-to) | 1375-1378 |
Number of pages | 4 |
Journal | IEEE Transactions on Magnetics |
Volume | 42 |
Issue number | 4 |
DOIs | |
State | Published - Apr 2006 |
Keywords
- Cochlear implant
- Electric field imaging
- Finite element method
- Tissue-interface model
- Volume conduction model